¤ıApplied model : B-530
¤ıCorrection time : 5sec cycle
¤ıMethod Duel detector is applied
|
¤ıApplied model : whole models
¤ıCorrection time : within 1 min
¤ıTolerance of data after re-correction : lower than 0.2%
¤ıCorrection method (process) : 1 point correction
¤ıCorrection Point : 0ppm (Nitrogen) correction or 400 ppm (natural standby) correction
400ppm correction (simple correction)
Convert the sample to the re-correction mode
Put the sample for natural standby.
After that, it is naturally re-corrected (1min), and ¡®Completed¡¯ lamp flickers after re-correction is completed.
0ppm correction (precision correction)
|
Convert the sample to the re-correction mode
Inject nitrogen gas (0ppm) after putting the sample into the re-correction chamber
After that, it is naturally re-corrected (1min), and ¡®Completed¡¯ lamp flickers after re-correction is completed.
|
ELT SENSOR¡¯s function approach to the correction of 99.8% from the 1-point returns to recalibrate.
|